Thigh-Worn Sensor for Measuring Initial and Final Contact During Gait in a Mobility Impaired Population: Validation Study
Johnson T; Kuntapun J; Childs C; Kerr A
Originally published in JMIR Biomed Eng · 2025-10-30
From the article
Selected source text. JournalOS removes markup and shortens excerpts where indicated. The full article provides the complete context.
Abstract
AbstractBackgroundAdapting physical activity monitors to detect gait events (ie, at initial and final contact) has the potential to build a more personalized approach to gait rehabilitation after stroke. Meeting laboratory standards for detecting these events in impaired populations is challenging, without resorting to a multisensor solution. The Teager-Kaiser energy operator (TKEO) estimates the instantaneous energy of a signal; its enhanced sensitivity has successfully detected gait events from the acceleration signals of individuals with impaired mobility, but has not been applied to stroke.ObjectiveThis study aimed to test the criterion validity of TKEO gait event detection (and derived spatiotemporal metrics) using data from thigh mounted physical activity monitors compared with concurrent 3D motion capture in chronic survivors…
Results
The Montreal cognitive assessment (MoCA) is a screening instrument that evaluates general mental capabilities, such as visuospatial abilities, executive functions, and orientation to time and space. The MoCA is rated between 1 and 30, with increasing score dictating better cognitive ability [31]. The functional ambulation category (FAC) evaluates walking ability in 6 levels, with a score of 0 defining no ability to walk, or requiring the help of 2 physiotherapist, and a score of 5 defining full capability to walk independently, including stairs [32]. The Rivermead mobility index (RMI) is an outcome measure used to assess mobility after stroke, rated between 0 and 15, with increasing score indicating better mobility [33] (Table 1). The AP4 (mean stance time=0.85) saw consistent underestimation in comparison to the ECMS (mean stance…
Discussion
This study aimed to test the concurrent validity of detecting gait events (ie, initial and final foot contact) and the derived metric—stance phase duration—from thigh acceleration data analyzed with the TKEO, compared with a gold standard 3D motion capture (Vicon) in chronic survivors of stroke. The excellent accuracy (bias=0.03 s) were similar to Flood et al (IC error=0.01s, FC error=0.02 s) [26]and Lozano-García et al (bias=0.08 s) [25] . In the context of a typical stance duration (nonparetic =1.01 s), the reported difference of 0.03 s (2 sampling frames) represents a 2.97% difference, with performance comparable to other commonly used methods of wearable sensor gait parameter measurement, collected by Pacini Panebianco [34] et al. The reported bias of 0.03 seconds could be considered clinically unimportant. A study by Wang [35] et…
Editorial reading guide
- How was movement measured, and what error or calibration information is reported?
- Can the finding transfer from the tested task and population to the intended sport setting?