Comparison of Echo Planar and Turbo Spin Echo Diffusion-Weighted Imaging in Intraoperative MRI.

James C Thorpe; Stefanie C Thust; Claire H M Gillon; Selene Rowe; Charlotte E Swain; Donald C MacArthur; Simon P Howarth; Shivaram Avula; Paul S Morgan; Rob A Dineen

Originally published in Journal of magnetic resonance imaging : JMRI · 2024-10-10

From the article

Selected source text. JournalOS removes markup and shortens excerpts where indicated. The full article provides the complete context.

Abstract

Background Diffusion-weighted imaging (DWI) is routinely used in brain tumor surgery guided by intraoperative MRI (IoMRI). However, conventional echo planar imaging DWI (EPI-DWI) is susceptible to distortion and artifacts that affect image quality. Turbo spin echo DWI (TSE-DWI) is an alternative technique with minimal spatial distortions that has the potential to be the radiologically preferred sequence. Purpose To compare via single- and multisequence assessment EPI-DWI and TSE-DWI in the IoMRI setting to determine whether there is a radiological preference for either sequence. Study type Retrospective. Population Thirty-four patients (22 female) aged 2-61 years (24 under 18 years) undergoing IoMRI during surgical resection of intracranial tumors. Field strength/sequence 3-T, EPI-DWI, and TSE-DWI. Assessment Patients were scanned with EPI- and TSE-DWI as part of the standard IoMRI scanning protocol. A single-sequence assessment of spatial distortion and image artifact was performed by three neuroradiologists blinded to the sequence type. Images were scored regarding distortion and artifacts, around and remote to the resection cavity.

Discussion

Overall the results from the blinded scoring and direct comparisons showed that for IoMRI, TSE‐DWI sequences had consistently superior image quality compared to EPI‐DWI sequences over a broad range of visual criteria. This can be explained by the severity of magnetic susceptibility artifacts introduced to the EPI‐DWI sequence by the presence of intracranial air during the surgery, despite surgical practice of filling the cavity with irrigation fluid and carrying out a temporary skin closure.

Conclusion

This study showed that when viewed in isolation, a TSE‐DWI sequence may have equal to or greater quality than an EPI‐DWI sequence in terms of distortion and signal pileup around and remote to the resection cavity, with greater visibility and confidence in reporting true diffusion abnormalities around the resection cavity for the TSE‐DWI sequence. In the multisequence comparison, a TSE‐DWI may be preferred overall for intraoperative assessments.