Directional deep brain stimulation of the subthalamic nucleus: A pilot study using a novel neurostimulation device.
Frank Steigerwald; Lorenz Müller; Silvia Johannes; Cordula Matthies; Jens Volkmann
Originally published in Movement disorders : official journal of the Movement Disorder Society · 2016-05-31
From the article
Selected source text. JournalOS removes markup and shortens excerpts where indicated. The full article provides the complete context.
Abstract
Introduction A novel neurostimulation system allows steering current in horizontal directions by combining segmented leads and multiple independent current control. The aim of this study was to evaluate directional DBS effects on parkinsonian motor features and adverse effects of subthalamic neurostimulation. Methods Seven PD patients implanted with the novel directional DBS system for bilateral subthalamic DBS underwent an extended monopolar review session during the first postoperative week, in which current thresholds were determined for rigidity control and stimulation-induced adverse effects using either directional or ring-mode settings. Results Effect or adverse effect thresholds were modified by directional settings for each of the 14 STN leads. Magnitude of change varied markedly between leads, as did orientation of optimal horizontal current steering. Conclusion Directional current steering through chronically implanted segmented electrodes is feasible, alters adverse effect and efficacy thresholds in a highly individual manner, and expands the therapeutic window in a monopolar review as compared to ring-mode DBS. © 2016 The Authors.
Results
We could determine efficacy thresholds for only 11 of 14 STNs, because a persistent microlesioning effect prevented reliable rigidity assessment in the others, whereas AE thresholds were determined for all 14 STNs. The AEs determining the upper limit of the TW were contractions of facial or hand muscles in 11 of 14 STNs, dysarthria in 6 of 14, and persisting dysesthesia in 1 of 14.
Discussion
Our results demonstrate, for the first time, the feasibility of horizontal current steering using a fully implanted neurostimulation device with directional leads and MICC technology. Despite the limitations of an acute monopolar review in the early postoperative period, they also provide first evidence of a beneficial impact of directional DBS on the TW.